Product Details

home // products // 3M Bair Hugger

3M Bair Hugger

3M™ Bair Hugger™ Therapy
Temperature Management Unit
Model 775
Quiet. Adjustable. Safe. Effective.
It’s everything you want from your
forced-air warming unit...and more.

3M is a trademark of 3M Company, used under license in Canada. BAIR HUGGER, BAIR PAWS and the BAIR HUGGER logo are trademarks of Arizant Healthcare Inc., used under license in Canada. Please recycle. Printed in USA. ©2011 Arizant Healthcare Inc. All rights reserved. 603032D 12/113 Forced-air warming was introduced to the world in 1987 with the launch of Bair Hugger therapy. Since that time, Bair Hugger warming units and blankets have safely and effectively warmed more than 135 million patients worldwide.

Our latest advancement in forced-air warming – the 3M™ Bair Hugger™ Model 775 temperature management unit – offers two airflow settings, hose-end temp sensing, precise temperature delivery and ease of use from a single, quiet warming unit. It’s also compatible with all Bair Hugger blanket styles – pediatric to adult – and 3M™ Bair Paws™ gowns.

Features
• Four temperature settings allow clinicians to deliver effective forced-air patient warming and cooling therapy. The ambient setting can complement your current patient cooling strategies.
• Three temperature sensors and adjustable airflow provide safe, quick temperature response.
• The Model 775 can track and report information on forced-air warming duration, over-temperature conditions and fault codes, all of which are accessible via the front panel – there’s no need to open the unit.
• Accessories for the Model 775 warming unit allow the unit to be mounted on a hard surface, an IV pole, bed rail or rolling cart.
Weight16 lb (7.3 kg)
Dimensions13
Operating TemperaturesHigh: 43° ± 1.5°C (109.4° ± 2.7°F) Med: 38° ± 1.5°C (100.4° ± 2.7°F) Low: 32° ± 1.5°C (89.6° ± 2.7°F) Ambient: (Room Temperature)
Fan Speeds and Relative NoiseHigh Fan Setting: 37 CFM (17.5 L/s) 53 dBA Low Fan Setting: 32 CFM (15.1 L/s) 48 dBA
Leakage CurrentMeets UL 60601-1 and IEC 60601-1 requirements. CAN/CSA-C22.2, No. 601.1
Safety SystemsAudible and visual alarms; Over-temperature ≤56°C, 53°C ± 3°C typical
FilterHigh-efficiency 0.2 μm filter
Device Ratings110-120 VAC, 50/60 Hz, 11.7 Amperes 220-240 VAC, 50/60 Hz, 7.2 Amperes 100 VAC, 50/60 Hz, 14.5 Amperes
CompatibilityUse only with Bair Hugger blankets and Bair Paws gowns
AccessoriesRolling cart, bed rail mount, accessory handle and teathered clip also available